  • Helmer i.series, 18 360137-1/D Undercounter Freezer Operation Manual 8 TechnicalSpecications Power Input voltage and frequency The requirements for a particular freezer are specied on the product specication label. The voltage tolerance is ±10% of the nominal voltage. Power consumption The power consumption for a particular freezer is specied on the product specication label. Power consumption is measured in full load Amperes. Input voltage Power consumption Circuit breakers 115 V, 60 Hz 5.75 A - 230 V, 50 Hz 2.8 A 6 A (quantity 2) 230 V, 60 Hz 3.1 A Load capacity for alarm contacts The terminals on the remote alar

  • Helmer i.series, 27 360137-1/D 10.6 Changing temperature controller setpoints The refrigerator is shipped from the factory with preset temperature setpoints. These setpoints are specic to the refrigerator’s intended use. These setpoints may be changed depending on organizational requirements. NOTE The Control Offset and Control Hysteresis are factory preset and should not be changed. 10.6.1 Monitor temperature offset value If the temperature displayed on the monitor does not match the actual chamber temperature, the setting for the Monitor Offset can be ch

  • Helmer i.series, 30 360137-1/D 12 TechnicalSpecications Power Input voltage and frequency The requirements for a particular freezer are specied on the product specication label. The voltage tolerance is ±10% of the nominal voltage. Power consumption The power consumption for a particular freezer is specied on the product specication label. Power consumption is measured in full load Amperes. Input voltage Power consumption Circuit breakers 115 V, 60 Hz 5.75 A - 230 V, 50 Hz 2.8 A 6 A (quantity 2) 230 V, 60 Hz 3.1 A Load capacity for alarm cont

  • Helmer i.series, 25 360137-1/D 9.3 Rear F G H I J K L A B C D E Rearfeatures(HPF105modelshown). Label Description Label Description A Insert for stacking bracket G External drain fan B Drain line H Rear panel C Compressor I Remote alarm interface D Product specication label J RS-232 COM port E Power connector K Power cord F Access port L Condensate evaporator Horizon Series™ Models: Components

  • Helmer i.series, 14 360137-1/D Undercounter Freezer Operation Manual 6 General Operation 6.1 Power on NOTE Allow unit to come to room temperature before power on. When the power is connected for the rst time, the freezer runs frequently to achieve normal operating temperature. This may cause an alarm to sound. When normal operating temperature is reached, the freezer runs normally and automatically clears the alarm. The freezer has a rechargeable backup battery that is switched off for shipping. Switch the battery ON to provide power to the moni

  • Helmer i.series, 17 360137-1/D i.Series™ Models: Maintenance Schedule 7 Maintenance Schedule Maintenance tasks should be completed according to the following schedule. Refer to the service manual and the i.C³ User Guide for more detail on the various tasks. NOTE These are recommended minimum requirements. Regulations for your organization or physical conditions at your organization may require maintenance items to be performed more frequently, or only by designated service personnel. Task Frequency Quarterly Annually As needed Test the high and low temperature alarms.  Test the power failure alarm (as required by

  • Helmer i.series, 16 360137-1/D Undercounter Freezer Operation Manual 6.7.3 Defrost status While the freezer is defrosting, the defrost icon appears on the Home screen. Homescreenwithdefrosticon. 6.8 Identifying active visual alarms If any alarms are active, the Alarm Condition indicator appears, with the type of alarm described below it. If multiple alarms are active, they are sequentially displayed for two seconds each, below the Alarm Condition indicator. If the alarm is for the chamber temperature, the display of the upper chamber temperature turns red. For more information about the i.C³ alarms, refer to the i.C³ User Guide. HomescreenwithanactiveHighTempera

  • Helmer i.series, 3 360137-1/D 1 About this Manual This chapter explains the symbols and conventions used in this manual, copyright information about this document, and trademark information for products supplied by Helmer. 1.1 Intended audience This manual is intended for use by end users of the freezer, and is to be used in conjunction with the i.C³ ™ User Guide, Undercounter Freezer Service Manual, Chart Recorder Operation Manual, and the Horizon Access Control Keypad User Guide, available on the CD shipped with the freezer. 1.2 Symbols an

  • Helmer i.series, 23 360137-1/D Horizon Series™ Models: Components 9 Components 9.1 Front and base FGHI A B C D E Frontandbasefeatures(HPF105modelshown). Label Description Label Description A Horizon control F Monitoring system battery backup B Door handle with lock G Alarm disable switch C Base H 6 A circuit breakers (230 V models only) D Leveling foot (casters are optional) I Main power switch E Chart recorder (standard on plasma storag

  • Helmer i.series, 15 360137-1/D i.Series™ Models: General Operation 6.6 Changing temperature setpoints The refrigerator is shipped from the factory with preset temperature setpoints. These setpoints are specic to the refrigerator’s intended use. Instructions for changing the temperature setpoints are outlined within Chapter 11: Alarm Settings, in the i.C³ User Guide. Refer to the i.C³ User Guide for instructions in changing temperature setpoints. 6.7 Understanding normal operation This sub-chapter describes some of the characteristics the freezer has during normal operation. 6.7.1 Understanding

  • 24 360137-1/D 9.2 Chamber A B E F G H CD Chamberfeatures(multiplemodelsshown). Label Description Label Description A Probe bottle E Unit cooler with fan guard B Shelf (laboratory/pharmacy models) F Chamber label C Roll out basket (optional) G Standard for adjusting storage components D Drawer (plasma storag

  • Helmer i.series, 26 360137-1/D 10 General Operation 10.1 Power on NOTE Allow unit to come to room temperature before power on. When the power is connected for the rst time, the freezer runs frequently to achieve normal operating temperature. This may cause an alarm to sound. When normal operating temperature is reached, the freezer runs normally and automatically clears the alarm. The freezer is shipped with the 9 V battery which is installed backwards. Reverse the battery installation and connect it to provide power to the monitoring system in the event of main power failure. 1 Plu

  • Helmer i.series, 6 360137-1/D 3.2 Preparing the temperature probe The temperature probe monitors chamber temperature. In addition to using standard probes installed by Helmer, external probes may be introduced through the existing rear port and immersed in the existing probe bottle. For the probe bottle, obtain: ► Approximately 4 oz (120 ml) of product simulation solution. Solution is a 1:1 ratio of water to propylene glycol (or equivalent low-temperature uid). Left:Probebottlewithtemperatureprobe.Right:Accessportonrearoffreezer. Tollatemperatureprobebottle ! CAUTION ► Clean bottle rst,

  • Helmer i.series, i 360137-1/D Section I: General Information ......................................1 1 About this Manual ........................................................3 2 Safety ..................................................................4 2.1 Labels ..........................................................................4 2.2 Avoidinginjury....................................................................4 3 Installation ..............................................................5 3.1 Locationrequirements..............................................................5 3.2 Preparingthetempe

  • 31 360137-1/D Size All dimensions are for the overall exterior and include items that protrude from the main unit. Width 24 in 610 mm Height (without casters or leveling feet) 34 in 864 mm Depth 28.5 in 724 mm Height notes: 1. With casters, the height of the undercounter freezer is approximately 36 in (920 mm). 2. With the leveling feet fully engaged (minimum height), the hei

  • Helmer i.series, 13 360137-1/D i.Series™ Models: Components 5.3 Rear F G H I J K L M A B C D E iPF105modelshownwithdetailfromvariousmodels Label Description Label Description A Insert for stacking bracket H Rear panel B Drain line I Remote alarm interface C Compressor J RJ-45 Ethernet port D Product specication label K USB port E Power connector L Power cord F Access port M Condensate evaporator G External drain fan Not shown RS-232 COM port (optional)

  • Helmer i.series, 28 360137-1/D Ifthislampisashing Then this setting is selected LOW TEMP and MONITOR Low Temp alarm setpoint MONITOR only Monitor Offset CONTROL only Control Offset CONTROL only Control Hysteresis NOTE The control lamp ashes when the Control Offset setting is selected. Press and release the SEL button to select the next setting (Control Hysteresis). The control lamp will continue to ash once, after the Control Hysteresis setting has been selected. 3 While pressing and holding the SET button, press and release the Up Arrow or Down Arrow button to change the value for the

Related Products and Documents (Freezer):

Comparable Devices:

# Manufacturer Model Document Type File Updated Pages Size
1 AHB System 8 Owner's handbook manual ahb/system-8-LJP.pdf 04 Apr 2024 92
2 LG RZ-23LZ41 Service manual lg/rz-23lz41-498.pdf 16 Aug 2022 36
3 Numatic LoLine Owner's instructions manual numatic/loline-5AJ.pdf 06 Sep 2023 16
4 Nipon Optics x Instruction manual nipon-optics/x-F3Y.pdf 22 Mar 2024 11
5 Hyundai H-VC1593 Instruction manual hyundai/h-vc1593-27U.pdf 20 Feb 2023 9
6 NEC SV-8100 Installation manual nec/sv-8100-89C.pdf 05 Aug 2023 56

Similar Resources:

  • AEG

    ARCTIS G 8 88 50 I

    (12 pages)
    ARCTIS G 8 88 50 iIntegrating FreezerOperating and Installation InstructionsAEG Hausgeräte GmbHPostfach 1036D-90327 Nürnberghttp://www.aeg.hausgeraete.de2222 107-63 …
  • Wine Guardian

    DS088

    (28 pages)
    Wine Guardian® Split Wine cellar Cooling System Models DS025, DS050, DS088, DS200 WGS40, WGS75, WGS100, WGS175 Quick Start Installaon Guide USA + 1 315-452-7400 | Europe + 41 52 224 0490 [email protected] Part # 15H0174-01 wineguardian.com …
  • Bosch

    KIL38A65

    (98 pages)
                  …
  • Perma Pure

    Baldwin series

    (44 pages)
    PPEERRMMAA PPUURREE Perma Pure LLC Tel: 732-244-0010 P.O. Box 2105, 8 Executive Drive Tel: 800-337-3762 (toll free US) Toms River, NJ 08754 Fax: 732-244-8140 www.permapure.com Email: [email protected] INSTRUCTION MANUAL CLASSIC THERM-ELECTRIC COOLER Classic Model 225CE …
  • LG

    LFC22760ST

    (35 pages)
    BOTTOM FREEZERREFRIGERATORPlease read this guide thoroughlybefore operating and keep it handy forreference at all times.P/No. 3828JL8075GModels / Modèles / ModelosCONGÉLATEURINFÉRIEURRÉFRIGÉRATEURVeuillez lire ce manuel en entier avant de mettreen marche l’unité et conservez-le à portée de lamain pour le consulter à tout moment.LFC22760**, LFD2286 …
  • Daewoo

    ERF-33.M

    (24 pages)
    INSTRUCTIONS MANUALENGLISH MODELS: ERF-334/5M.. ERF-336/7M.. ERF-34*M/A.. ERF- …

Comments, Questions and Opinions: