  • MELAG MELAtherm 10, Accessories for interior cleaning 56 When using filter elements, comply with the following:  Hollow-bodied instruments with an interior diameter ≤ 0.8 mm require filter elements.  When using the Basis basket with injector rail (incl. 2 central filters), the filter disc must be removed from the universal adapter.  Comply with the regular replacement intervals for the filter inserts (see section Filter inserts). Table 2: Silicone inserts for universal adapter Qty. Silicone inse

  • Accessories for exterior cleaning 12 Basis basket with injector rail (incl. 11 nozzles and clamp springs) Art. no. 00197 Intended use If hollow-bodied instruments are to be rinsed use the basis basket with injector rail (incl. 11 injector nozzles and clamp springs). The hollow-bodied instruments (e.g. dental surgical aspirator tips) are stuck on the injector nozzles and

  • MELAG MELAtherm 10, Accessories for exterior cleaning 24 Used in conjunction with  Flex basket 1, art. no. 80010  Flex basket 2, art. no. 80020  Flex basket 3, art. no. 80030  Ear speculum attachment Flex 1, art. no. 80070, 80080,  Ear speculum attachment Flex 2, art. no. 80070, 80090,  Ear speculum attachment Flex 3, art. no. 80070, 20100,  Insert rack for impression trays, scissors and clamps, art. no. 80110  Insert rack for hinged instruments, art. no. 80120  Injectior basket Flex 1, art. no. 80740 Flex basket 6 Art. no. 80255

  • MELAG MELAtherm 10, Accessories for exterior cleaning 16 Malfunctions and their remedy Perform all the following measures before contacting the service dept. Malfunctions Remedy The central filter is new, but the pin does not spring out of the control display. 1. Check whether the central filter has been fitted correctly. 2. Check the loading of the injector rail. Seal the unused connections with sealing screws (art. no. 80140). 3. Always place instruments on the adapter. If possible, seal non- used adapters with the suitable silicone sealing cap. If the two

  • MELAG MELAtherm 10, Accessories for interior cleaning 42 Clamp spring for injector nozzle Art. no. 00196 Intended use The clamp spring facilitates the fixing of light hollow-bodied instruments onto the injector nozzle so that this does not slip through the rinse pressure. It is used exclusively with the injector nozzle. Application The clamp spring is pressed together lightly and slid over the injector nozzle. Used in conjunction with  Injector nozzle, art. no. 73860 Irrigation sleeve for instruments with Ø 2-11 mm Art. no. 80260 Intended use The irrigation sleeve acts as an adapter for the treatment of hollow-b

  • MELAG MELAtherm 10, Accessories for interior cleaning 58 Used in conjunction with  Universal adapter (incl. 3 inserts and filter disc), art. no. 73904  Filter disc housing (incl. ceramic filter disc), art. no. 73905  Triple distributor (incl. ceramic filter disc), art. no. 73903 Adapter for exterior spray channels Art. no. 74135 Intended use The adapter serves the rinsing of exterior (spray) channels in transfer instruments and other non- standardized hollow-bodied instruments with small external diameter. Application The adapter for external spray channels is screwed onto a connection of the injector ra

  • MELAG MELAtherm 10, Accessories for exterior cleaning 25 PLEASE NOTE Avoid improper loading combinations such as multiple stacking of the instruments. Otherwise, the cleaning performance could be compromised. In accordance with national regulations regarding the cleaning test, the following limitations apply to the maximum number of lying hinged instruments per flex basket:  Flex basket 6: 20 pieces As part of validation, adaptations of program parameters may be necessary. PLEASE NOTE In order to optimise the cleaning result, decontaminate hinged instrumen

  • MELAG MELAtherm 10, Accessories for exterior cleaning 37 Dividing insert for sieve cassette Art. no. 00191 Intended use The dividing insert facilitates the stability of the instruments in the sieve cassette so as to avoid spray shadow. Application The dividing inserts are inserted into the sieve cassette lengthways. The instruments are divided lengthways across the resulting three segments (see figure on the right). It is intended that the instruments are able to move. Used in conjunction with  Sieve cassette, art. no. 00185 Protrusion guard for the sieve cassette Ar

  • MELAG MELAtherm 10, Accessories for interior cleaning 66 Silicone sealing cap, blue (10 pcs.) Art. no. 89061 Intended use If an instrument is not fitted to the adapter, the adapter can be sealed quickly and easily with the silicone sealing cap. Application Slide the silicone cap as far onto the adapter until it is tight. To enable easy removal, do not slide on the silicone sealing cap to the fullest extent. Sliding the seal cap on and off in a dry state can be difficult Used in conjunction with  Connection for hoses with interior Ø 6mm, e

  • MELAG MELAtherm 10, Contents Connection for hoses with interior Ø 6mm, internal thread ....................................................................... 70 Silicon hose Ø 10/6 mm, running metre ................... 70 Hose extension with screw connections ................... 71 Examples for basic configurations .................. 72 Basic configuration for the ENT practice ...................... 72 Basic configuration for the gynaecological practice ..... 72 Basic configuration for the dental practice ................... 73 Basic configuration for the OMF surgery ..................... 73

  • Accessories for interior cleaning 46 Adapters for tips Adapter for tips M3.0 x 0.5 mm, external thread Adapter for tips M3.6xPH1.5 P0.5, internal thread Adapter for tips M3.0 x 0.35 mm, external thread Art. no. 80750 Art. no. 80751 Art. no. 80752 Compatible with: KaVo SONOsoft, PiezoLUX EMS Piezon Swiss Instruments W&H Piezo Scaler: Tigon, Tigon

  • MELAG MELAtherm 10, Examples for basic configurations 72 Examples for basic configurations Basic configuration for the ENT practice Basis basket with injector rail (incl. 11 blind screws), art. no. 00200 | Flex basket 2, art. no. 80020 | Flex basket 3, art. no.80030 | Instrument basket compact, art. no. 00195 | Ear specula attachment Flex 2, art. no. 80090 | Top frame for nasal specula (incl. 2 fixing clamps), art. no. 80435 Basic configuration for the gynaecological practice Basis basket without injector rail, art. no. 00188 | Flex basket 6, art. no. 80255| Flex basket specula, art. no. 80410.

  • MELAG MELAtherm 10, Accessories for exterior cleaning 19 Insert rack for 5 trays / 10 half-trays Art. no. 80590 Space requirements Intended use The tray mount accommodates instrument trays (LxWxH) 24 x 18 x 1.7 cm to max. 29 x 19 x 2.0 cm. It is not stackable. Application Remove the coarse soiling from the trays before treatment in the device. Always place multiple trays in the tray mount with the impression pointing in the same direction. Depending on the size, 5 - 10 instrument trays can be positioned (see Fig. 10). Check the instrument trays for soiling after treatm

  • MELAG MELAtherm 10, Principles of the load configuration 9 Principles of the load configuration The 6 segment principle All accessories are placed in the basis basket with or without an injector rail. The surface of the basis basket is divided into six segments to ensure the optimal use of space. The size of all insert racks, instrument baskets, washing baskets and flex baskets correspond to one or more segments of the basis basket. To this end, the next pages provide one of the following diagrams for each accessory.

  • Accessories for exterior cleaning 36 Sieve cassettes and inserts Sieve cassette Art. no. 00185 (sieve cassettes), Art. no. 80185 (sieve cassett with dividing insert and protrusion guard) Intended use The sieve cassettes facilitate the treatment of instrument sets or larger quantities of individual instruments in combination with the insert rack

  • MELAG MELAtherm 10, Accessories for exterior cleaning 23 Stackable insert racks and baskets (flex system) The flex baskets can be subject to variable combinations and stacked. Page 10 lists one of a number of many combinations in the flex baskets. Flex basket 1 Flex basket 2 Flex basket 3 Art. no. 80010 Art. no. 80020 Art. no. 80030 Space requirements Space requirements Space requirements Intended use The flex basket facilitates the treatment of instruments lying flat e.g. tweezers, mirrors, scissors etc. Application The flex basket is

  • MELAG MELAtherm 10, General instructions for preparation and treatment 6 General instructions for preparation and treatment 1 Hollow-bodied instruments 1.1 The automatic decontamination of hollow-bodied instruments requires that they be arranged in the washer-disinfector on adapters on the injector rail or via suitable load elements to ensure sufficient throughflow. Universal dental aspirator tips with 11 mm and 16 mm connections can be treated in instrument baskets in a standing position. The distal end must point upwards. This must be considered separately during validation. 1.2 After decontamination, complex hollow bodie

  • MELAG MELAtherm 10, Optimizing the cleaning performance and value-retention of the instruments 8 3.5 Treat only those instruments in the washer-disinfector which have been expressly authorized for this purpose by the manufacturer. As a rule, these instruments are marked with this symbol, either directly or in the treatment instructions. 3.6 If required, order the treatment instructions from the respective manufacturer in accordance with EN ISO 17664. 3.7 Comply with the treatment instructions provided by the instrument manufacturer, especially those pertaining to computability with process agents. 3.8 Do not trea

  • Accessories for interior cleaning 67 Sealing screw for injector rail and distributor Art. no. 80140 Intended use The sealing screw is used to seal non-required connections on the injector rail or the distributors. Application The sealing screw is screwed into a non-used connection. Seal (male) for Luer-Lock Art. no. 80170 Intended use The seal (male)

  • MELAG MELAtherm 10, Accessories for interior cleaning 51 Used in conjunction with  Silicone sealing cap, blue (10 pcs.), art. no. 89061  Filter disc housing (incl. ceramic filter disc), art. no. 73905  Basis basket with injector rail (incl. 2 central filters), art. no. 80440 Adapter for the heads of the KaVo handpieces Art. no. 80630 Intended use This adapter serves the acceptance of removable KaVo handpiece heads. Application The adapter is screwed onto a connection on the injector rail with central filter, a triple distributor or a single filter housing (injector rail without central filter) and tighten

Related Products and Documents (Washer):

Comparable Devices:

# Manufacturer Model Document Type File Updated Pages Size
1 GE moisture.IQ Operation & user’s manual ge/moistureiq-9XJ.pdf 07 Nov 2022 178
2 Blackberry Curve Getting started manual blackberry/curve-731.pdf 03 Sep 2023 58
3 Maytag RTS1700CAL Manual  maytag/rts1700cal-27O.pdf 24 Apr 2023 18
4 Rosewill RC-GS1024 Operation & user’s manual rosewill/rc-gs1024-XAD.pdf 28 Nov 2023 14
5 AEG Milano 40 Operation & user’s manual aeg/milano-40-MI8.pdf 17 Nov 2023 32 0.22 Mb
6 Avtronix AD-808Q Operation & user’s manual avtronix/ad-808q-3QS.pdf 07 Aug 2022 41

Similar Resources:

  • AEG

    Lavamat 74900

    (48 pages)
    603095OFFCOLDAQUA ALARMENERGY SAVINGEASY IRONREFRESH 3040COLD40 - 60 MIXDEL. RINSESDRAINSPIN40403030DELICATE SPIN14001000700500PREWASHMAIN WASHRINSERINSESPINENDOVERDOSINGDOORDELICATESWOOLCOTTONS74900LINEN/EASY-CARES40SILK 30RINSE HOLDTIMESAVERWASH TIME8 KgRINSEDELAYTIMERSTARTPAU SEPRE-WASHSTAIN SENSITIVELAVAMAT 74900The environmentally friendly washing machi …
  • Beko

    WTV 7732 XS0

    (76 pages)
    2820525485_EN / 02-08-17.(11:08)Washing MachineUser’s ManualWTV 7732 XS0EN / CZ / SK / SLDocument Number=Plně automatická pračkaUživatel ManuálníAutomatická pračkaPoužívateľská príručkaPralni strojUporabniški priročnik …
  • Kenmore

    796.4127

    (32 pages)
    Model/Modelo: 796.4127#Kenmore Elite®Front-LoadingAutomatic WasherLavadora automática de carga frontalP/N MFL67360435Sears Brand Management CorporationHoman Estates, IL 60179 U.S.A.www.kenmore.comwww.sears.comUse & Care GuideManual de Uso y CuidadoEnglish / Español # = color number, numero de colorPara obtener los mejores resultados, deberá utiliz …
  • Frigidaire

    Full Size Tumble Action Washers

    (12 pages)
    InstallationInstructionsFull Size Tumble Action WashersBefore beginning installation, carefully read these instructions. This will simplify theinstallation and ensure the washer is installed correctly and safely. Leave these instructionsnear the washer after installation for future reference.NOTE: The electrical service to the washer must conform with local …
  • CUDA

    2412

    (38 pages)
    2412AUTOMATIC PARTS WASHERFor technical assistance or the Kärcher dealer nearest you, visit our website at www.CudaUSA.com■ 1.043-356.0 ■ 1.043-379.0 ■ 1.043-357.0OPERATING INSTRUCTIONSMAINTENANCE INSTRUCTIONSTURNTABLEOIL SKIMMERON89143840-19.801-125.08.914-384.0 …
  • LG

    F1K2CS2T

    (101 pages)
    DWG.No.TRIG.DR COMPLETED도면 DR 완료완료일 : LG 전자 (주)LG Electronics Inc.UNIT SCALENSmmREVIEWEDDESIGNEDRELATED DWG.CHECKED APPROVED16.04.01MODELINGITLETAPPROVEDPREPAREDDATEREF. NO.REV. NO. REVISION DESCRIPTION1234RELEASING THIS DRAWING WITHOUT PERMISSION LGElectronics SHOULD BEACCUSED ACCORDING TO THELAWS AND COMPANY RULES.이 도면은 L …

Comments, Questions and Opinions: