  • Instruction 0011-9026 Page 21 accuracy don’t forget to saturate fi lter saturator wool packing as such gas standards are usually supplied “bone dry”. 6.2.2 O 2 FYRITE Fluid Only Fresh FYRITE fl uid will absorb all O 2 from approximately 100 samples containing 10% O 2 . After completing test (as outlined in Section 3.0 Steps 1 through 9) O 2 FY-

  • Bacharach FYRITE Gas Analyzer CO2 and O2 Indicators, Instruction 0011-9026Page 24 mine replacement parts necessary for repair, or drain instrument and return if factory service or repair is desired. 6.5 Cleaning FYRITE Use only soapy lukewarm water if cleaning is required (lukewarm water is usually suffi cient). NOTE: Use of gasoline, naptha, carbon tetrachloride or any other organic solvent or oil will destroy plastic and rubber parts. 6.6 Replacing FYRITE Plastic and Rubber Parts (Refer to Illustrated Parts List FYRITE CO 2 /O 2 [Section 7.0 and Fig. 33]) Replace plastic parts when cracked or crazed in location exposed to fl uid and rubber par

  • Bacharach FYRITE Gas Analyzer CO2 and O2 Indicators, Instruction 0011-9026Page 14 Proper O 2 content for any fuel fi red is the lowest O 2 value that will ensure complete, clean combustion with adequate safety margin for variations in fuel, draft, atmospher- ic conditions, and mechanical wear. Consult manufacturer of heating equipment or local authorities for specifi c recommendations. Figure 15 illustrates the relationship between CO 2 , excess air and oxygen for gas and oil. For guidance, it should be added that modern boilers are capable of 80% or greater combustion effi ciency. Rough guidelines for setting conventional power burners are 10-15% excess air for nat

  • Bacharach FYRITE Gas Analyzer CO2 and O2 Indicators, WARRANTY Bacharach, Inc. warrants to Buyer that at the time of delivery this Product will be free from defects in mater ial and manufac ture and will conform substan tially to Bacharach Inc.’s ap plicable specifi cations. Bacharach’s liability and Buyer’s remedy under this warranty are limited to the repair or replacement, at Bacha- rach’s option, of this Product or parts thereof returned to Seller at the factory of manufac ture and shown to Bacharach Inc.’s reasonable satisfaction to have been defective; provided that written notice of the defect shall have been given by Buyer to Bacharach Inc. within one (1) year after the date of delivery of this Produc

  • Bacharach FYRITE Gas Analyzer CO2 and O2 Indicators, Instruction 0011-9026 Page 5 Figure 5 Figure 5a Figure 5b Figure 6 CO 2 FYRITE OPERATION O 2 FYRITE OPERATION (RED FLUID) (BLUE FLUID) 5. Holding FYRITE upright (Fig. 5), loosen locknut at rear of scale. Slide scale (Fig. 5a) until top of fl uid column lines up with zero line on scale (Fig. 5b). Tighten scale locknut. NOTE: When setting scale zero, hold FYRITE vertically as shown and level with eyes while sighting across scale to top

  • Bacharach FYRITE Gas Analyzer CO2 and O2 Indicators, Instruction 0011-9026 Page 1 INSTRUCTIONS 0011-9026 OPERATION/MAINTENANCE BACHARACH FYRITE MODELS CO 2 and O 2 INDICATORS 1.0 DESCRIPTION (Refer to Figure 1) The FYRITE employs the well-known “Orsat” method of volumetric analysis involving chemical absorption of a sample gas, such as carbon dioxide or oxygen. The reagent used to absorb carbon dioxide (CO 2 ) is potassium hydroxide (dyed red), and chromous chloride (blue) is the absorbent for oxygen (O 2 ). The unique feature of the FYRITE is th

  • Bacharach FYRITE Gas Analyzer CO2 and O2 Indicators, Instruction 0011-9026 Page 9 Figure 12. When plunger valve is depressed, a gas sample is pumped through top reservoir with center bore is sealed off. Figure 13. When plunger valve is released, the gas sample is locked into FY- RITE and the top reservoir is opened to center bore so that gas sample can pass through absorbing fl uid. Figure 14. Absorption of gas sample by fl uid cre- ates suction, which causes diaphragm to fl ex up and fl uid to rise in center bore to replace gas absorbed. 4.0 PRINCIPLE OF OPERATION (Refer to F

  • Bacharach FYRITE Gas Analyzer CO2 and O2 Indicators, Instruction 0011-9026 Page 23 20 16 12 8 4 %C O 2 0 20 16 12 8 4 %CO 2 0 ® 20 16 12 8 4 %CO 2 0 ® GASKET BODY TOP CAP ASSEMBLY Top Gasket used in 7%, 20% & 21% Models only ® 20 16 12 8 4 %CO 2 0 Figure 21a. Alternate fi lling meth- od. Invert FYRITE (with bottle in place) to upright position. Figure 22. Examining top gasket for warpage. Figure 23. FYRITE inverted in test disk (overnight) for leakage test. Figure 24. Top gasket properly centered in top fl ange of body.

  • Instruction 0011-9026Page 10 To make a test with the FYRITE, the metal Sampling Tube at one end of Sampling Assembly Hose is inserted into the gas to be analyzed. The rubber Connector Plug at the other end of the Rubber Hose is then fi rmly pressed down on the spring-loaded Valve of the instrument (See Figure 12). This simultaneously opens a passage into the Top Reservoir and seals off

  • Bacharach FYRITE Gas Analyzer CO2 and O2 Indicators, Instruction 0011-9026Page 12 0-7.6% CO 2 or O 2 CO 2 tests of controlled atmospheres in fruit, vegetable, and meat storage rooms. Oxygen determination in fl ammable gases. Oxygen tests to check atmospheres made inert with nitrogen (silos, fuel tanks, etc.) 0-20% CO 2 or 0-21% O 2 Flue gas combustion tests, oxygen defi ciency tests, and CO 2 tests of heat treating atmospheres. Checking oxygen concentrations in hydrogen-cooled generators and oil-sealed inert-gas transformers. 0-60% CO 2 or O 2 Checking CO 2 in inert gas blankets in tankers and barges carrying gasoline and other com bustibles. CO 2 tests on lime kilns. Checking CO 2 in sewage

  • Bacharach FYRITE Gas Analyzer CO2 and O2 Indicators, Instruction 0011-9026 Page 7 10. This step completes CO 2 or O 2 FYRITE gas sample analysis. A few points to re- member when reading the FYRITE: (a) FYRITE accuracy is within ±1/2% CO 2 or O 2 compared to actual value. (b) Always handle FYRITE by fi ns to ensure body heat is not absorbed by fl uid. (c) A delay in reading of 5 or 10 seconds may decrease accuracy of reading slightly but longer delays may cause substantial error. (d) The FYRITE is calibrated to indicate on a “dry” basis for fl ue gas samples, which are normally fully saturated with moisture. Failure to artifi cially moisten wool packing in Filter Saturator Tube when actual sample is n

  • Bacharach FYRITE Gas Analyzer CO2 and O2 Indicators, Instruction 0011-9026 Page 13 Excess combustion air is heated and carries some of this heat to the fl ue where it is wasted. The FYRITE CO 2 Analyzer is used to adjust combustion excess air to a mini- mum (maximum CO 2 ), which will permit clean effi cient combustion. Calculation of combustion effi ciency is possible (assuming com plete combustion) if percentage of CO 2 and net temperature of combustion products are known. 5.4.3 What is Proper CO 2 Value? Proper CO 2 is that which will ensure complete, clean combustion with some safety margin for variations in fuel, draft, atmospheric con ditions, and mechanical wear. Consult manufacturer of heating

  • Bacharach FYRITE Gas Analyzer CO2 and O2 Indicators, Instruction 0011-9026Page 26 with fi nger and squeeze Aspirator Bulb (Figure 27). Bulb should remain fi rm. If Bulb collapses, check Bulb and Hose to Connector Tip for cracks or other source of leakage. Replace defective parts. If there are no apparent leaks in the Sampling Assembly, replace Inlet Check Valve. Small hole end of Inlet Check Valve fi ts in Hose and large hole end fi ts in the Aspirator Bulb (Figure 26). Now seal end of metal Sampling Tube with fi nger, and collapse Aspirator Bulb. (Fig- ure 28). If the Bulb returns to origin

  • Instruction 0011-9026 Page 11 FYRITE CO 2 refi ll fl uid may develop a white insoluble precipitate upon aging. The precipitate does not affect fl uid performance in any way, but if present, should be fi ltered out to prevent adherence to internal surfaces of the FYRITE. To fi lter, place a clean piece of thin cloth or netting loosely over the open CO 2 FYRITE

  • Bacharach FYRITE Gas Analyzer CO2 and O2 Indicators, Instruction 0011-9026Page 18 The gas concentration read on the FYRITE is directly dependent upon the mass of air in the sample. The aspirator bulb used in the FYRITE is a constant-volume pump, not a constant-mass pump. Altitude, therefore, affects the FYRITE reading due to the air’s density changing with altitude, thus requiring higher CO 2 or O 2 concentrations to reach the same mark on the scale. Use the following table to fi nd the altitude correction (e.g., add the correction to the reading to get the correct concentration). TABLE 5-1. ALTITUDE CORRECTION

  • Bacharach FYRITE Gas Analyzer CO2 and O2 Indicators, Instruction 0011-9026Page 32 Current Former Part # List # Description # Req'd 11-0152 10-0019 Connector Tip with Tube 1 11-0156 10-0029 Rubber Tubing, 10' Length 1 11-0118 10-0020 Rubber Tubing, 6" Length 2 11-0165 10-0030 Filter Tube, Aluminum 1 19-5004 19-9004 Gas Collecting Bladder with Orifi ce 1 11-0120 10-0022 Aspirator Bulb 1 11-0138 10-0023 Inlet/Outlet Valve (Red) 2 11-01

  • Instruction 0011-9026Page 2 (b) FYRITE Fluid Strength To check fl uid strength. NOTE: When repeating procedures as outlined in Section 3.0 Steps 7 through 9 (absorbing and reading percent O 2 and ab- sorbing and reading percent CO 2 ) and before venting FYRITE to atmosphere for next sample, if the reading increases by more than 1/2 percent for either CO 2 or

  • Bacharach FYRITE Gas Analyzer CO2 and O2 Indicators, FYRITE SAFETY PRECAUTIONS The operator (s) of this instrument should thoroughly familiarize themselves with the applicable safety precautions before handling or using the FYRITE for gas analysis. Review Figures 1a,1b, 1c and 1d on this page. Make certain to follow the steps out- lined below and read the fl uid handling precautions for your personal safety. 1) Always use protective equipment such as safety goggles, gloves, and protective clothing as illustrated in Figure 1a. 2) Read the instruction manual thoroughly. When fi lling the FYRITE with the appli- cable fl uid charge, use safety equipment and perform the operation in the vicinity of running tap water (Figure 1b). 3) The F

  • Bacharach FYRITE Gas Analyzer CO2 and O2 Indicators, Instruction 0011-9026Page 16 As a rule of thumb, background gases/vapors may be present in con centrations up to 1/2% by volume (5000 ppm) before they present a signifi cant interference problem to the CO 2 or O 2 FYRITE (20/21% ranges). An exception exists with the action of ammonia on Oxygen FYRITE Fluid. Ammonia neutralizes the acidic solution of chromous chloride, and therefore use of the O 2 FY- RITE with even trace amounts of ammonia in the backgrounds is not advised unless suitable fi lters are used. Inert gases, even in concentrations approaching 100% by volume, do not interfere with either the CO 2 or O 2 FYRITE. In general, interferences can be of a

  • Bacharach FYRITE Gas Analyzer CO2 and O2 Indicators, TABLE OF CONTENTS 1.0 DESCRIPTION................................................................................................................................................1 2.0 FYRITE INSPECTION BEFORE AND DURING TEST ..............................................................................1 2.1 Pre-Operational Check .......................................................................................................................1 2.2 FYRITE FLUID HANDLING PRECAUTIONS ......

  • Bacharach FYRITE Gas Analyzer CO2 and O2 Indicators, Instruction 0011-9026Page 36 Bacharach, Inc. 621 Hunt Valley Circle, New Kensington, PA 15068 Phone: 724-334-5000 • Fax: 724-334-5001 • Toll Free: 800-736-4666 Web: www.mybacharach.com • E-mail: [email protected]

  • Bacharach FYRITE Gas Analyzer CO2 and O2 Indicators, Instruction 0011-9026 Page 31 0011-0106 0011-0118 0011-0118 0011-0122 (0011-0180 + 0011-0130) 0011-0119 0011-0138 0011-0138 0011-0120 0011-0152 7.2 PARTS LIST FOR FYRITE SAMPLING ASSEMBLIES STANDARD SAMPLING ASSEMBLY Part No. 11-7029 Gases saturated with water vapor (combustion products). Dry gases when fi lter material is wetted.

Related Products and Documents (Security System):

Comparable Devices:

# Manufacturer Model Document Type File Updated Pages Size
1 Zanussi Electrolux Z 57/3 W Instruction booklet zanussi-electrolux/z-57-3-w-5H3.pdf 02 Oct 2022 20 0.49 Mb
2 KETRON Audya Operation & user’s manual ketron/audya-2Y7.pdf 16 Apr 2023 162
3 Sharp LC-60LE925X Operation manual sharp/lc-60le925x-C3M.pdf 16 Sep 2022 96
4 IBM SAN768B Installation, service and user manual ibm/san768b-A4A.pdf 20 Feb 2023 161
5 Electrolux 65320 K User instructions electrolux/65320-k-8Q4.pdf 06 Jul 2022 40 0.29 Mb
6 Cooper Lighting Metalux BA SERIES Specification sheet cooper-lighting/metalux-ba-series-845.pdf 10 Feb 2024 1

Similar Resources:

  • Fortin

    EVO-ONE

    (6 pages)
    ONEREV.: 20170323ADDENDUM - SUGGESTED WIRING CONFIGURATION ADDENDA - SCHÉMA DE BRANCHEMENT SUGGÉRÉVehicle DATA functions supported (Functional if equipped)Fonctions du véhicules supportés en DATA (Fonctionnelle si est équipé)VEHICLEVEHICULESYEARS ANNÉESLockUnlockArmDisarmTachometerTachometerDoor StatusTrunk StatusHand-Brake StatusFoot-Brake StatusRAM …
  • Visonic

    AMBER SELECT

    (1 pages)
    D-301439 1 AMBER SELECT / AMBER SELECTX Setting Up Your System Quick Start Guide To set up your system, follow the steps below. For more detailed information, see the AMBER SELECT / AMBER SELECTX User Guide. Select a central location in your home for the AMBER SELECT / AMBER SELECTX where you spend most of your time. The AMBER SELECT / AMBER SELECTX can …
  • Networker

    TS590

    (40 pages)
    TS590Intruder Alarm Control PanelOperators Manual1ADCB237654890ENTESC_~SYSTEM OPEN17:30 01 JanSIL EN TZONE OMITCHIMENE W COD E 2 4 H r OMI TWALK TESTBELLTESTPar t Se tPar t Se tPart Se tRE S E T1ABCD234567ENT ESC809Sett ing th e SystemEnter your pass code XXXXthen leave the protected area.Unsetting the SystemGo directly to the keypad anden ter y ou r p assc …
  • LJ&L

    CamoCam IV

    (10 pages)
    CamoCam™ Remote Video Surveillance CamoCam IV™ L J & L Products, Inc. CAMOCAM™ REMOTE VIDEO SURVEILLANCE Features • Portable • Lightweight • Versatile • Open Architecture User’s Guide …
  • Cisco

    Nexus 7000 Series

    (438 pages)
    Cisco Systems, Inc.www.cisco.comCisco has more than 200 offices worldwide. Addresses, phone numbers, and fax numbers are listed on the Cisco website at www.cisco.com/go/offices.Cisco Nexus 7000 Series NX-OS MPLS Command ReferenceDecember 2016Text Part Number: OL-24994-02 …
  • stabo

    freecomm 650 Set

    (36 pages)
    stabo freecomm 650 SetPMR 446 Handsprechfunkgerät / radio handheldBedienungs-anleitungOwnersManualWichtig/ImportantLesen Sie vor Inbetriebnahme des Geräts alle Bedienhinweise aufmerksamund vollständig durch. Bewahren Sie diese Anleitung sorgfältig auf,sie enthält wichtige Betriebshinweise.Before your installation please read this manual carefullyand kee …

Comments, Questions and Opinions: