Text of Thermo Scientific DLT User Guide:

  • Thermo Scientific DLT, Thermo Scientific Harris Ultra-Low Temperature Freezers Installation and Operation 4 5 Operation (Harris DLT and SLT Freezers) 5.1 Control Panel Features Figure 3. Harris DLT and SLT Control Panel with Optional Backup System OFF EXTREME AMBIENT ALERT CLEAN FILTER POWER FAILURE TEMPERATURE FAILURE ALARM TEST ALARM RESET CABINET TEMPERATURE CONTROL SET POINT ALARM BATTERY LOW VOLTAGE LOW WARM ALARM SET POINT COLD AL ARM SET POINT POWER ON ALARM ON 1 11 2 7 8 9 10 17 18 193 SUPPLY ON SUPPLY EMPTY BATTERY LOW BACKUP SYSTEM 45 6 20 21 12 16 15 1413 SET POINTS LOCKED 9. Alarm Battery Low Indicator. This indicator lights when the charge on the alarm battery is low. P

  • Thermo Scientific DLT, 9 Thermo Scientific Harris Ultra-Low Temperature Freezers Installation and Operation 7 Operation (ELT Freezers) 7.1 Control Panel Features Before the initial start up, take some time to become familiar with the controls on your freezer. Figure 6 illustrates the Harris ELT Series Chest and Upright Freezer control panel. Figure 6. Harris ELT Series Control Panel 1. Keyed Power On/Alarm On switch. 2. LED digital temperature display. 3. Recessed temperature control set point adjustment. 4. Temperature set point adjust butt

  • Thermo Scientific DLT, Thermo Scientific Harris Ultra-Low Temperature Freezers Installation and Operation 13 Temperature Recorders All six inch recorders use pressure-sensitive chart paper (1 box @ 50 charts included); no inking is required. Chart Paper Special Voltages Standard voltages are listed under Voltage in Specification Charts associated with each product category. Standard voltages are available at no charge; alternative voltages may be available at an extra charge. All voltages must be specified when ordering. Contact Sales for more information. Electronic Backup Systems CO 2 Backup System. Choose built-in or free-standing. Injects

  • Thermo Scientific DLT, Thermo Scientific Harris Ultra-Low Temperature Freezers Installation and Operation 14 Surge Protection Surge Protector. Provides protection against external electrical spikes (temporary high voltage condition). Built-in surge protection is optional on all freezers models; free-standing surge protection is also optional on all models. Remote Alarm/Monitoring RS-232 Data Port and Software Extended Warranty Options Note: In addition to the standard 12-Month full warranty on the complete product (

  • Thermo Scientific DLT, Thermo Scientific Harris Ultra-Low Temperature Freezers Installation and Operation 6 5.2.5 Alarm Test (Harris DLT and SLT Freezers) To test the cold and warm alarms, complete the following steps: 1. Press and hold the Alarm Test pad. The digital temperature display indicates rising temperature. When the temperature reaches the warm alarm value, the alarm sounds. Note: The temperature of the refrigerated space does not change during this procedure. Only the sensor is heated. 2. Release the Alarm Test pad, the sensor and display retu

  • Thermo Scientific DLT, Thermo Scientific Harris Ultra-Low Temperature Freezers Installation and Operation 1 1 Safety Considerations WARNING! Do not modify or change system components. Replacement parts must be O.E.M. exact replacement equipment. Modification or use of the equipment in a manner other than expressly intended may cause death or serious injury. This includes use of user-supplied components and materials not specifically designed for the unit. Reconfiguring the controller may cause death or serious injury. The manufacturer shall not be liable for any damages, including incidental and/or consequential damages, regardle

  • Thermo Scientific Harris Ultra-Low Temperature Freezers Installation and Operation i Table of Contents 1 Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 2 General Recommendations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

  • Thermo Scientific DLT, Important For your future reference and when contacting the factory, please have the following information readily available: Model Number: Serial Number: Date Purchased: The above information can be found on the dataplate attached to the equipment. If available, please provide the date purchased, the source of purchase (manufacturer or specific agent/rep organization), and purchase order number. IF YOU NEED ASSISTANCE: SALES DIVISION Phone: 828/658-2711 800/221-4201 FAX: 828/658-0383 LABORATORY PARTS and SERVICE

  • Thermo Scientific DLT, Thermo Scientific Harris Ultra-Low Temperature Freezers Installation and Operation 5 5.2 Start Up Refer to Section 5.1 and Figure 3 on page 4 as you complete the following procedures. 5.2.1 Turning the Power On To start up the Harris DLT and SLT Series Freezers, complete the following steps: 1. Plug the freezer into the power outlet (refer to Section 4.2 on page 2). 2. Turn the key switch to the POWER ON position. The Power On LED and the Cabinet Temperature LED illuminate. Note: The audible alarm function is not active at this time. Refer to Section 5.2.3 for information about setting the alarms. In addition, the Set Points will not be lo

  • Thermo Scientific DLT, Thermo Scientific Harris Ultra-Low Temperature Freezers Installation and Operation 10 8 Maintenance and Troubleshooting WARNING! Unauthorized repair of your freezer will invalidate your warranty. Contact Technical Service at 1-800-438-4851 for additional information. CAUTION! Maintenance should only be performed by trained personnel. 8.1 Condenser Maintenance 8.1.1 Cleaning the Condenser Clean the condenser at least every six months; more often if the laboratory area is extremely dust prone. To clean the condenser, complete the following steps: 1. Pull the grill open. 2. Remove the filter. Check

  • Thermo Scientific DLT, © 2007 Thermo Fisher Scientific. All rights reserved. “Suva ® ” is a registered trademark of DuPont. Armaflex ® is a registered trademark of Armacell International GmBH. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. 2007 Thermo Fisher Scientific. All rights reserved. “Suva ® ” is a registered trademark of DuPont. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.

  • Thermo Scientific DLT, 8 Thermo Scientific Harris Ultra-Low Temperature Freezers Installation and Operation Table 1. Backup System Flow Rates Lbs/hr @ 75°F Ambient Temperature (Flow Rate) Lbs/hr @ 75°F Ambient Temperature (Flow Rate) Center Air Temperature -50°C-60°C-70°C Center Air Temperature -60°C-70°C-80°C Liquid CO 2 Liquid LN 2 Chest Freezers Chest Freezers 3 ft 3 Empty 2.832.963.1 3 ft 3 Empty 2.7 3.1 3.5 1/2 Full 2.4 2.7 3.1 1/2 Full 2.2 2.6 2.9 Full 1.8 2.1 2.4 Full 1.7 1.9 2.2 7 ft 3 Empty 4.1 4.7 5.5 7 ft 3 Empty 3.9 4.5 4.1 1/2 Full 3.5 4.0 4.6 1/2 Full 3.3 3.8 4.3 Full 2

  • Thermo Scientific Harris Ultra-Low Temperature Freezers Installation and Operation 11 9 Optional Equipment The following equipment is available for some freezer models either as factory-installed or as field-installed options. Refer to the Accessories section on page 11 for more information on these and other options, including inventory racks and storage boxes. For additional informat

  • Thermo Scientific DLT, Thermo Scientific Harris Ultra-Low Temperature Freezers Installation and Operation 7 6.3 Operation To activate the optional backup system, complete the following steps: 1. Push the Power Switch to the ON position. The digital temperature display window shows the current cabinet temperature. 2. Press and hold the Push to Set button to display the set point temperature in the digital temperature display window. 3. Use a small screwdriver to rotate the Set Point Adjust screw until the correct backup operating temperature shows in the display window. The recommended set point is 10 to 15°C above normal cabin

  • Thermo Scientific DLT, 15 Thermo Scientific Harris Ultra-Low Temperature Freezers Installation and Operation WEEE Compliance WEEE Compliance. This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96EC. It is marked with the following symbol. Thermo Fisher Scientific has contracted with one or more recycling/disposal companies in each EU Member State, and this product should be disposed of or recycled through them. Further information on our compliance with these Directives, the recyclers in your country, and information on Thermo Scientific products which may assist the detection of substances subject to the RoHS Directive are

  • Thermo Scientific Harris Ultra-Low Temperature Freezers Installation and Operation 12 Inventory Component Description, Ultra-Low Temperature Freezers Description Catalog No. For All Models Fiberboard boxes and grid dividers are packaged in lots of one dozen Fiberboard box, 5-1/4 in. square x 2 in. high, one dozen. 5954 Fiberboard box, 5-1/4 in. square x 3 in. high, one dozen. 5956 Grid d

  • Thermo Scientific DLT, Thermo Scientific Harris Ultra-Low Temperature Freezers Installation and Operation ii 7 Operation (Harris ELT Freezers) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 7.1 Control Panel Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9 7.2 Start Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9 7.2.1 Turning the Power On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Related Products and Documents (Freezer):

Comparable Devices:

# Manufacturer Model Document Type File Updated Pages Size
1 Gemini XLB Installation and operation manual gemini/xlb-Q6F.pdf 23 Nov 2023 19
2 Carrier 48N Installation instructions manual carrier/48n-6U8.pdf 18 Jul 2022 24 2.21 Mb
3 Ningbo BPS 101 Operating manual ningbo/bps-101-87S.pdf 16 Oct 2023 92
4 Sony CDX-L410X - Fm/am Compact Disc Player Operating instructions manual sony/cdx-l410x-fm-am-compact-disc-player-2BT.pdf 18 Mar 2024 18 0.55 Mb
5 Ericsson Orion Service manual ericsson/orion-I16.pdf 22 Dec 2023 24
6 Agri-Fab H.D. Lawn Roller 45-02163 Owner's manual agri-fab/h-d-lawn-roller-45-02163-FXP.pdf 01 Oct 2023 2 0.18 Mb

Similar Resources:

  • WHISPER KOOL

    MINI PLATINUM SPLIT VERSION 3

    (29 pages)
    MINI PLATINUM SPLITVERSION 3OWNER’S MANUALThe Coolest Thing In Wine StorageNOTICE: To activate the split system warranty, the installing certied HVAC/R service tech must complete the split system warranty checklist and send back to CellarCool. …
  • True Fitness

    TWT-27F

    (2 pages)
    Specifi cations subject to change without notice.Chart dimensions rounded up to the nearest 1/8" (millimeters rounded up to next whole number).ROUGH-IN DATASt. Charles Industrial Center • P.O. Box 970 • O’Fallon, Missouri 63366(636)240-2400 • FAX (636)272-2408 • (800)325-6152 • www.truemfg.comParts Dept. (800)424-TRUE • Parts Dept. FAX# ( …
  • Bosch

    B30IR900SP

    (57 pages)
    en Instruction manual .............................................................. 2fr Notice d'utilisation ............................................................ 19es Instrucciones de uso ........................................................ 38RefrigeratorRéfrigérateurFrigoríficoB30IR900SP …
  • Huawei

    NetCol5000-A050

    (27 pages)
    1Overview1Copyright © Huawei Technologies Co., Ltd. 2020 All rights reserved.Issue: 05Code: 31500BLXDate: 2020-07-08NetCol5000-A050 In-row Air Cooled Smart Cooling Product Quick GuideCategoryProduct ModelDimensions Without Packing (mm) (H x W x D) Indoor UnitNetCol5000-A0502000 x 600 x 1200Outdoor UnitNetCol500-A060(52273539)1107 (677) x 1356 (1217) x 1 …
  • IKEA

    604.621.54

    (77 pages)
    设 计 DESIGN 校 对 CO.BY 蔡宇超图号/物料编码 PART NO. 标准化 STANDARD 图纸名称 PART NAME 业务 seller 图样标记 REL 版本号REV.NO. 重量(g) WEIGHT 比例scale材料 MATERIAL 80g双胶纸批 准 AUTHORIZE共 页 TOTAL SHEETS 第 页 NO. OF SHEETS美的冰箱事业部 MIDEA REFRIGERATION DIVISION 技术要求 标记 MARKER 1:1变� …
  • Hussmann

    C6X-LEP

    (9 pages)
    BDCDDCAEFP/N 0479366_EC6X-LEPTechnical Data SheetP/N 0479366_ENSF®CertifiedMarch 2008We reserve the right to change or revise specificationsand product design in connection with any feature ofour products. Such changes do not entitle the buyer tocorresponding changes, improvements, additions orreplacements for equipment previously sold or shipped.Item Par …

Comments, Questions and Opinions: