  • Kinesis QTUG, 16 4.1 QTUG test requirements To perform a QTUG test on a patient, ensure the patient is wearing comfortable walking shoe. You will require the following to complete a QTUG test: 4. Set up the QTUG test 4m (13’1”) of linear floor space. One standard chair with seat approx. 46 cm (18”) high and total chair height approx. 65 cm (25.6”) high with arm-rests. 4m / 13’1” Ensure the chair used for the test is stable (i.e. chair without wheels).

  • 31 10. Parameter definition Definition of mobility paramters produced by QTUG. Parameter Description Falls risk estimate (%) Statistical risk of falling (defined for community dwelling older adults over 60 years of age) TUG recording time (s) Time to complete TUG test Temporal gait parameters Parameter Description Time taken to stand (s) Time from ‘Go’ to first heel strike or toe-off

  • Kinesis QTUG, 29 Sensors At the end of the assessment, remove the sensors from the elasticated bandages. The sensors can be wiped down using standard clinical alcohol wipes to reduce risk of infection. After cleaning, insert the sensors into the sensor dock to be fully charged for the next assessment. Please turn off the sensors after they are fully charged. Tablet At the end of the assessment, press the HOME button on the tablet to close the QTUG application. Please turn off the tablet after use. If the battery power is low, connect the tablet to the USB power adaptor for charging. Before cleaning, it is essential to remove the connecting USB cable for the tablet from the power inlet so

  • Kinesis QTUG, 36 Specifications Model Shimmer2R w/450mAH Battery Power supply Input 100 - 240V AC 50-60Hz Output 7.5 VDC 15W nominal, 1.7A Max Current consumption 700mA (7.5V input) Weight < 700 grams (1.4 pound) Operating Conditions +5 ˚C – +40 ˚C (20% – 95% Relative Humidity) Storage/Transport Conditions -20 ˚C – +60 ˚C (20% – 95% Relative Humidity) Bluetooth Radio Transmit Band 2.4Ghz Modulation GFSK, DQPSK, and 8DPSK Frequency Range 2400MHz – 2483.5MHz Output Power Min: 5 dBm; Typical: 6 dBm; Max: 6 dBm Bluetooth Radio Receiver Bandwidth 75kHz Frequency Range 2400MHz – 248

  • Kinesis QTUG, 15 Turning the sensor on and off directly Use a paper clip or a pen, press the restart pin to turn on the sensor. The activity LED turns orange. Wait a moment for the sensor to boot up. When it is ready, the activity LED turns green. To turn off the sensor, press and hold the restart pin for 7 seconds until the activity LED turns off. 3.5 Resetting the sensor Use a paper clip or a pen to reset the sensor. Press the restart pin once and the activity LED turns orange. When it is ready, the activity LED turns green. To reset the sensor with the sensor dock, press the RESET but

  • Kinesis QTUG, 19 6 Enter DOB, height and weight by tapping the - and + buttons or the number and entering manually. Select a gender by tapping the MALE or FEMALE button. Tap ‘New Test’ to proceed. 5 Type in a new patient ID to create a new patient profile and tap the SUBMIT button to proceed to the patient profile page. Do not use a patients’ Personally Identifiable Information (PII) as part of patient ID. Enter height and weight to nearest cm or ft/inch and kg or lbs.

  • Kinesis QTUG, 35 Maximum forward rotation speed x Height (deg.m/s) Related to maximum linear velocity of shank in forward direction Average forward rotation speed x Height (deg.m/s) Related to minimum linear velocity of shank in forward direction Minimum side-to-side rotation speed x Height (deg.m/s) Related to minimum linear velocity of shank in side-to-side direction Maximum side-to-side rotation speed x Height (deg.m/s) Related to maximum linear velocity of shank in side-to-side direction Average side-to-side rot

  • Kinesis QTUG, 3 This User Guide introduces you to QTUG (Quantitative Timed Up and Go), and helps you quickly learn how to perform a mobility assement or a falls risk assessment on a patient. For detailed information about the QTUG application, refer to the application help-files, which you access through the QTUG Help Menu. The User Guide also contains useful information about configuring QTUG, transferring data files from the tablet to a PC or microSD card for organizing your QTUG patient files, and using all of the QTUG features. Intended

  • Kinesis QTUG, 33 Turn parameters Parameter Description Pre-turn time (s) Time from ‘go’ to median gait event of TUG test Post-turn time (s) Time from median gait event of TUG to end of test Ratio of pre-turn to post-turn times Ratio of Time from ‘go’ to median gait event of TUG to Time from median event of TUG to end of test Time taken to turn (s) Time taken to turn Number of strides in turn Number of steps in turn Turn steps/time ratio Ratio of the number of steps taken to turn to the time taken to turn Angular velocity parameters Parameter D

  • Kinesis QTUG, 8 QTUG package includes the following components. 1. What’s in the box? 7” tablet Power adaptor USB cable 2 wireless inertial sensors Sensor dock Elasticated bandages Tablet and its accessories Sensor and its accessories Miscellaneous Measuring tape Carrying case Quick Start Guide Warranty and Safety Health Technologies Improving lives through Innovation QTUG Quantitative Timed Up and Go NovaUCD, Belfield Innovation Park, Belfield, Dublin 4, Ireland | www.kinesis.ie, [email protected] Health Technologies QTUG WARRANTY AND SAFETY © Kinesis Health Technologies 2014 All Rights Reserved Health Technologies QTUG WARRANTY AND SAFETY © Kinesis Healt

  • Kinesis QTUG, 24 6 When the patient is reseated, tap the STOP button to stop recording data. The sensor data is displayed on the screen. 7 On the application, tap the ACCEPT button to confirm that the test is valid (i.e. the patient performed the test correctly), and to save data. Tap RETEST if the test is invalid and you wish to perform test again and discard data, the start screen is displayed.

  • Kinesis QTUG, Address: NovaUCD Belfield Innovation Park Belfield Dublin 4, Ireland Internet address: http://www.kinesis.ie Email address: [email protected] © 2014 Kinesis Health Technologies Ltd., All Rights Reserved. 2

  • Kinesis QTUG, 4 Read these safety messages carefully. This device contains an RF transmitter. It is also an intentional RF receiver and even if other equipment complies with CISPR emissions requirements, those devices may interfere with the operation of this device. Radio Information Transmit Characteristics: 2.4GHz Bluetooth radio using GFSK, DQPSK, and 8DPSK modulation and 75kHz bandwidth. Frequency Range in MHz: 2400-2483.5. Output Power in dBm: 5-6. This equipment has been tested and found to comply with the EMC limits for the Medical Device Directive 93/42/EEC (EN 55011 Class A and EN 6

  • 17 4.2 Physical set up 2 Position the chair at the start of the 3m (9’10”) space. 1 Clearly mark the start and end (turn) points on the floor placed exactly 3m (9’10”) apart, measured from patient’s toes when seated. 3m / 9’10” 3m / 9’10” end (turn) point start point Ensure the 3m (9’10”) distance is accurately measured from the chair to the end point. Incorrectly me

  • Kinesis QTUG, 22 3 The sensors must be placed in the correct orientation with the socket facing down and the QTUG logo facing out, the Red sensor on the Right leg and the bLue sensor on the Left leg. QTUG QTUG user’s point of view socketsocket QTUG R UP R UP QTUG L UP L UP 4 When the patient is ready, say ‘Go’ and tap the START button on the QTUG application. When START is tapped, the activity LED on the sensor flashes alternately green-orange in streaming mode. Failure to use the correct orientation will result

  • 30 9. Troubleshooting If there are any problems which cannot be rectified immediately, contact Kinesis technical support. Problem Cause Remedy The tablet does not turn on The battery is flat Check that tablet is fully charged. The sensor does not connect The sensor is turned off Check that sensor is switched on. The sensor does not turn on The battery is flat Check that sensor is full

  • Kinesis QTUG, 12 3.1 Sensor layout 3. Getting started with the sensor QTUG R UP L UP BT RADIO ID Kinesis Health Technologies Ltd. Model: Shimmer2R w/450mAH Battery Contains: FCC ID: X2W-SR7-1; T9-RN42 IC: 8838A-SR71; 6514A-RN42 MEDICAL EQUIPMENT INTERMITTENT OPERATION WITH RESPECT TO ELECTRIC SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL60601-1, IEC/EC60601, IEC60601-1-2 Made In Ireland 06/14 activity LED restart pin charging socket R indicates the Red sensor attached to the Right leg L indicates the bLue sensor attache

  • Kinesis QTUG, 23 5 The patient must get up from the chair, walks 3m (9’10”) to the turning point as marked on the floor, turn at the turning point, walk back to the chair and sit down. 3m / 9’10” 3m / 9’10” The patient has to walk exactly 6m (19’7”). The test is calibrated to give valid results based on accurate measurement of the 3m (19’7”) distance from the chair to the turning point. Incorrectly measured distance may result in incorrectly derived gait and mobility parameters.

  • Kinesis QTUG, 5 • Reorient or relocate the receiving device. • Increase the separation between the equipment. • Connect the equipment into an outlet on a circuit different from that to which the other device(s) is connected. • Consult the manufacturer for help. The use of portable and mobile RF equipment may have an impact on this and other pieces of medical equipment. The use of accessories, cables, or transducers other than those specified in this manual can significantly increase emissions performance and degrade immunity performance of the product. Also, by using an accessory, transducer, or cable with the product, other than th

  • Kinesis QTUG, 28 Erase patient data from internal storage Prior to erasing patient data, ensure that data are backed up to a PC or laptop. To erase the data from the internal storage on the tablet, tap on ‘My Files’ and tick the ‘Kinesis’ folder. On top of the screen, tap on the bin icon and confirm with ‘Yes’ to delete the ‘Kinesis’ folder. Permanently erased data cannot be recovered. Erase patient data from microSD card, only if a microSD card is used To erase the data from the microSD card, tap on ‘My Files’ / ‘extSdCard’ and tick the ‘Kinesis’ folder. On top of the screen, tap o

  • Kinesis QTUG, 34 Average side-to-side rotation speed (deg/s) Average angular velocity in the side-to-side direction during the assessment Minimum forward rotation speed (deg/s) Minimum forward angular velocity in the sagittal plane during the assessment Maximum forward rotation speed (deg/s) Maximum forward angular velocity during the assessment Average forward rotation speed (deg/s) Average forward angular velocity during the assessment Variation in forward rotation speed (%) Coefficient of variation in forward angular

  • Kinesis QTUG, 13 3.2 Sensor dock layout micro USB power connector Kinesis Model: QTUG Charger Revision: 1.0.0 Use only supplied power adapter. Approved for use only with Kinesis QTUG system. For indoor use only RESET button socket power LED charging LED

  • 20 7 If the sensors are not connected yet, the application will take approx. 5 seconds to automatically connect to the sensors. 8 When sensors have connected successfully, the QTUG test screen will appear.

  • Kinesis QTUG, 10 2.2 Charging the battery When the battery power is low, the tablet emits a warning tone and displays a ‘low battery’ power message. If the battery is completely discharged, the tablet cannot be turned on immediately when the charger is connected. Allow a depleted battery to charge for a few minutes before running on the tablet. Charging with the charger Connect the large end of the USB cable to the USB power adaptor and then plug the end of the USB cable into the mulitpurpose jack under the tablet. Ensure the tablet is charged before performing a QTUG test. After fully charging, unplug the USB cable from the tablet, and then unplug it from t

  • Kinesis QTUG, 14 3.3 Charging the sensor Place the sensor in the sensor dock, ensuring sensor socket connected to socket in sensor dock. The power LED on the sensor dock turns only green when the sensor is turned on, and when the sensor and the sensor dock is successfully connected. The sensor battery is not fully charged while the charging LED remains orange. If the charging LED turns off, the sensor battery is fully charged. The sensor battery should last approximately 5 hours under normal working conditions. 3.4 Turning the sensor on and off Turning on and off with the sensor dock To turn on the sensor, place the sensor in the sensor

  • Kinesis QTUG, 25 An accepted and valid test displays and saves the test results. The definitions of these results are explained in the table below. Definition Description TUG test time Time taken to complete the TUG test. Falls risk estimate (FRE) Statistical risk of falls based on a model derived from community dwelling older adults. Comparison to ref. data Comparison of each patients’ mobility (as measured by QTUG) when compared to average values for the population. Parameters deemed out of the normal range are shown here (if no paramete

Related Products and Documents (Tablet):

Comparable Devices:

# Manufacturer Model Document Type File Updated Pages Size
1 Envision G2219 Operation & user’s manual envision/g2219-45Y.pdf 11 Jul 2022 24 0.61 Mb
2 Napoleon PT600RB-INFRARE Operation & user’s manual napoleon/pt600rb-infrare-A54.pdf 21 Dec 2023 35
3 Citizen REF CH-491E Instruction manual citizen/ref-ch-491e-IIB.pdf 10 Oct 2022 10 0.27 Mb
4 SEW MOVIAXIS MX Operating instructions manual sew/moviaxis-mx-3G2.pdf 24 Mar 2024 212
5 LG AX500 Red Datasheet lg/ax500-red-M53.pdf 08 May 2023 2
6 ASROCK ALIVENF5-VSTA R2.0 Installation manual asrock/alivenf5-vsta-r2-0-2FD.pdf 08 Apr 2023 26 1.22 Mb

Similar Resources:

  • Dell

    Inspiron 510M

    (2 pages)
    www.dell.com | support.dell.com1. network connector2. modem connector3. S-video TV-out connector4. USB connectors5. AC adapter6. AC adapter connector7. IEEE 1394 connector8. PC Card slot9. infrared sensor10. audio connectors11. power button12. touch pad13. touch pad buttons14. display latch15. module bay1. connecteur réseau2. connecteur modem3. connecteur d …
  • BungBungame

    KALOS 2

    (28 pages)
    52各種お問い合わせ窓口● 電話:0120-160-067● 受付時間:平日11:00~18:00● 電子メール:[email protected]● Webサイト:www.bungbungame.com洽詢方式● 電話:0809-080-566● 電子郵件:[email protected]● 營業時間:周一~周五10:00~18:00(UTC+8)� …
  • Samsung

    Galaxy S4

    (217 pages)
    GH68_XXXXXA Printed in KoreaANDROID SMARTPHONEUser ManualPlease read this manual before operating yourphone and keep it for future reference. This document is watermarked because certain important information in the document has not yet been verified.This document has been sent to you as a draft and for informational purposes only.The document should o …
  • Dell

    Inspiron 20

    (2 pages)
    Install the standInstallez le socleSet up the keyboard and mouseInstallez le clavier et la souris1 2 5 Finish Windows setupTerminez l’installation de WindowsSee the documentation that shipped with the keyboard andmouse.Reportez-vous à la documentation fournie avec le clavier et la souris.Connect the power adapterBranchez l’adaptateur secteurPress the …
  • AIPTEK

    MyNote Premium II

    (244 pages)
    English/Deutsch/Français/Italiano/Español/Nederlands/Português/日本語/한국어/简体中文/繁體中文 …
  • HP

    Laserjet,Color Laserjet 4550

    (302 pages)
    HP Color LaserJet CM3530 MFP SeriesUser GuideProduct use Product managementProduct maintenanceProblem solvingAdditional product information:www.hp.com/support/cljcm3530mfpwww.hp.com/go/usemyMFP …

Comments, Questions and Opinions: